Sensoray 118 Bedienungsanleitung Seite 38

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 38
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 37
Sensoray Model 118 Smart A/D™ Instruction Manual 36 Specifications
9.2 Sensor Specifications
Notes
1. Measurement resolution and accuracy is specified for the default (versus
high-speed) measurement mode. Derate resolution and accuracy by 75%
when using the high-speed measurement mode.
2. Channels configured for 4-20mA current loops return data values
representing a percentage of full-scale current (4mA= 0%,
20mA=100%). For example, a 12mA current would produce the data
value 50.0, indicating 50% of full-scale.
3. This is the default sensor type after a board reset. It is provided for
back-compatibility with earlier Sensoray Smart A/D products. Use of
this type is not recommended for new designs, as higher resolution is
available on the 5V, 200uV resolution range.
4. Declaring a channel’s sensor type as Disabled removes the channel from
the scan loop. Data values returned from disabled channels are
indeterminate and should be ignored.
5. Gage specifications assume 4-wire bridge circuit, 120 minimum
impedance, 10V excitation applied by sensor channel.
6. Use the following function to express gage measurement resolution in
load units:
For example, take the case of a pressure gage that is rated at 3mV/V at
100 PSI full scale. The Smart A/D would be able to resolve a load to:
7. SDC is the Sensor Definition Code that is used by the SetSensorType
command to declare a channel’s sensor type. See Section 6.2.1 for
details.
Table 9: Sensor Specifications
Sensor Type
Measurement
SDC
7
(hex)
Range Resolution
1
Accuracy
1
Thermocouple B 0 to +1820°C 0.1°C 3.3°C 24
C 0 to +1820°C 0.1°C 2.1°C 23
E -270 to +990°C 0.1°C 0.8°C 01
J -210 to +760°C 0.1°C 0.6°C 1B
K -270 to +1360°C 0.1°C 1.0°C 1C
N -270 to +1347°C 0.1°C 0.9°C 22
T -270 to +400°C 0.1°C 0.6°C 1D
S 0 to +1760°C 0.1°C 3.0°C 1E
R 0 to +1760°C 0.1°C 2.8°C 1F
Thermistor Omega 44006 or 44031 -55°C to +145°C 0.01°C 0.05°C 1A
RTD Cu 10, 0.0367/°C 0 to +119°C 0.1°C 0.6°C 2C
Pt 100, 0.385/°C -200 to +800°C 0.05°C 0.2°C 18
Pt 100, 0.392/°C -200 to +800°C 0.05°C 0.2°C 19
Ni 200, 1.10/°C -58.9 to +151.6°C 0.02°C 0.07°C 28
Ni 1K, 5.60/°C -50.0 to +129°C 0.025°C 0.08°C 29
Gage
5
-500 to +500 mV 5µV
6
30µV 0F
DC Voltage -5 to +5V
3
500µV 600µV 00
-5 to +5V 200µV 600µV 15
-500 to +500mV 20µV 40µV 16
-100 to +100mV 5µV 30µV 17
Current Loop
2
4 to 20mA 0.01% 0.08% 11
Resistance 0 to 400 0.02 0.04 0A
0 to 3K 0.125 0.25 14
0 to 600K 31 130 20
Disabled
4
Not Applicable — sensor removed from scan loop 13
Resolution
RatedLoad
210
6
×()RatedVoltageOut×
---------------------------------------------------------------------------=
Resolution
100PSI
210
6
×()
0.003
V×
----------------------------------------------0.017PSI==
Seitenansicht 37
1 2 ... 33 34 35 36 37 38

Kommentare zu diesen Handbüchern